3-76
Basic Operations
TOS9200/9201
3.17
Initialization
•
When the tester is initialized, the contents of panel memory and all programs are
cleared. Before initializing the tester
, mak
e sure these contents will not be needed.
T
o initialize the tester
, press the SHIFT k
e
y + the PO
WER switch to start it up. Ini
-
tialization will start.
All contents of memory are deleted, and the tester returns to the
f
actory settings. (Hold the SHIFT k
e
y do
wn until “KIKUSUI ELECTR
ONICS
CORP
.
” disappears.)
■
F
actor
y settings
•
T
est-Condition Settings screen (A
CW)
T
est v
oltage (V
OL
T
A
GE)
: 0.00 kV
Frequenc
y (FREQ)
: 50 Hz
Lo
wer judgement (LO
WER)
: OFF
Lo
wer current (LO
WER)
: 0.10 mA
Upper current (UPPER)
: 0.20 mA
Of
fset (OFFSET)
: OFF
T
imer (TIMER)
: ON
T
imer set v
alue (TIMER)
: 0.5 s
Start v
oltage (ST
AR
T)
: 0 %
V
oltage rise time (RISE
TIME)
: 0.1 s
V
oltage f
all time (F
ALL
TIME)
: 0.0 s
V
oltage range (V RANGE)
:
A
UT
O
Response
fi
lter (RESPONSE)
: SLO
W
LO
W/GU
ARD of the GND (GND)
: GND
Scanner channel
:
ALL OPEN
CONT
A
CT CHECK
: OFF
•
T
est-Condition Settings screen (DCW)
T
est v
oltage (V
OL
T
A
GE)
: 0.00 kV
Lo
wer judgement (LO
WER)
: OFF
Lo
wer current (LO
WER)
: 0.10 mA
Upper current (UPPER)
: 0.20 mA
T
imer (TIMER)
: ON
T
imer set v
alue (TIMER)
: 0.5 s
Start v
oltage (ST
AR
T)
: 0 %
V
oltage rise time (RISE
TIME)
: 0.1 s
Judgement w
ait time (W
AIT
TIME)
: 0.3 s
LO
W/GU
ARD of the GND (GND)
: GND
Scanner channel
:
ALL OPEN
CONT
A
CT CHECK
: OFF
NOTE
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......