TOS7200
55
Specifications
General specifications
Environment
Installation location
Indoors and at altitudes up to 2000 m
Warranty
range
Temperature
5 °C to 35 °C (41 °F to 95 °F)
Humidity
20 %rh to 80 %rh (no condensation)
Operating
range
Temperature
0 °C to 40 °C (32 °F to 104 °F)
Humidity
20 %rh to 80 %rh (no condensation)
Storage
range
Temperature
-20 °C to 70 °C (-4 °F to 158 °F)
Humidity
90 %rh or less (no condensation)
Power requirements
Nominal voltage range
(allowable voltage range)
100 Vac to 240 Vac
(85 Vac to 250 Vac)
Power
consumption
At rated load
30 VA maximum
Allowable frequency range
47 Hz to 63 Hz
Insulation resistance
30 MΩ or more (500 Vdc) [between AC LINE and the chassis]
Withstand voltage
1390 V AC for 2 seconds, 10 mA or less [AC LINE to chassis]
Earth continuity
25 Aac/0.1 Ω or less
Safety
*1
Conforms to the requirements of the following directive and standard.
Low Voltage Directive 2014/35/EU
EN 61010-1 (Class I
*2
, Pollution degree 2
*3
)
Electromagnetic compatibility
(EMC)
*1
Conforms to the requirements of the following directive and standard.
EMC Directive 2014/30/EU
EN 61326-1 (Class A
*4
)
EN 55011 (Class A
*4
, Group 1
*5
)
EN 61000-3-2
EN 61000-3-3
Under following conditions
1. Used HV test leadwires TL08-TOS which is supplied.
2. No discharge occurs at outside of the tester.
3. Used the shielded cable which length is less than three meters
when the SIGNAL I/O is used.
Dimensions
See Dimensions.
Weight
Approx. 2 kg (4.4 lb)
Accessories
AC power cord
1 pc.
TL08-TOS high-voltage
test leadwires
1 set
Setup guide
1 pc.
Quick reference
English: 1pc., Japanese: 1pc.
Safety information
1 pc.
CD-ROM
1 pc.
*1.
Only on models that have CE marking on the panel. Not applicable to custom order models.
*2.
This instrument is a Class I equipment. Be sure to ground the protective conductor terminal of the instrument.
The safety of the instrument is not guaranteed unless the instrument is grounded properly.
*3.
Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric
strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive pollution will occur except
for an occasional temporary conductivity caused by condensation
*4.
This is a Class A equipment. This product is intended for use in an industrial environment. This product may cause
interference if used in residential areas. Such use must be avoided unless the user takes special measures to
reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
*5.
This is a Group I equipment. This product does not generate and/or use intentionally radio-frequency energy, in
the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or
inspection/analysis purpose.