TOS6200A/6210
Basic Operation
4-15
Changing the test current during the test:
You can change the test current with the rotary knob during a test. Note that the
value should fall within the upper and lower limits.
The test current cannot be changed if the key lock function is activated.
• Changing the test current when the resistance value is close to the upper reference
value may precipitate a FAIL status.
Switching the screen during the test:
During testing, you can change the screen display as follows using the F5 (DISP)
key.
To stop the test:
To stop the test, press the STOP switch.
4.3.2
Ending the Test
Test will be terminated if any of the following conditions is established:
•
A test time has elapsed (with the timer ON) [PASS judgment].
•
A measured value out of the range of the set reference values has been detected
[FAIL judgment].
•
The STOP switch was pressed.
•
The tester has entered protection status.
■
PASS judgment
When the lower limit judgment function is ON and the measured value falls
between the upper and lower reference values, the tester returns a PASS judgment.
When the lower limit judgment function is OFF, the tester returns a PASS judgment
for the measured value if the value is less than the upper reference value.
When a PASS judgment is returned for the measured value, "PASS" appears in the
upper right part of the LCD, the PASS LED indicator lights, and the buzzer sounds.
The PASS judgment is generally displayed for about 0.2 seconds (default value).
You can set the PASS judgment display time in the range 0.2 s to 10.0 s or to HOLD.
For information on altering this setting, see 4.4, "System Setup.”
NOTE
DISP
1.4
s
MAIN
MAIN
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
TEST
50Hz
3.00
V
25.0
A
0.120
NORM
DISP
TEST
3.00
V
25.0
A
1.4
s
0.120
NORM
Press
the F5 key.
Standard display
Large numbers display
1.4
s
MAIN
MAIN
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
PASS
50Hz
3.00
V
25.0
A
0.120
NORM
Summary of Contents for TOS6200A
Page 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Page 14: ...1 8 General TOS6200A 6210...
Page 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Page 34: ...3 8 Part Names and Functions TOS6200A 6210...
Page 76: ...4 42 Basic Operation TOS6200A 6210...
Page 134: ...7 6 Maintenance TOS6200A 6210...
Page 150: ...A 8 Appendix TOS6200A 6210...