2-4
Installation and Preparation for Use
TOS6200A/6210
2.3
Connecting the AC Power Cord
• This product is a piece of equipment that conforms to IEC Safety Class I
(equipment that has a protective conductor terminal). Be sure to earth
ground the product to prevent electric shock.
The product is grounded through the power cord ground wire. Connect the
protective conductor terminal to earth ground.
• Use the supplied power cord to connect to the AC line.
If the supplied power cord cannot be used because the rated voltage or the plug
shape is incompatible, have a qualified engineer replace it with an appropriate
power cord that is 3 m or less in length. If obtaining a power cord is difficult, con-
tact your Kikusui agent or distributor.
• The power cord with a plug can be used to disconnect the product from the AC
line in an emergency. Connect the plug to an easily accessible power outlet so that
the plug can be removed from the outlet at any time. Be sure to provide adequate
clearance around the power outlet.
•
Do not use the supplied power cord for other devices.
This product is a piece of equipment that conforms to IEC Overvoltage Category II
(energy-consuming equipment that is supplied from a fixed installation).
Procedure
1.
Check that the POWER switch is turned off.
2.
Check whether the AC power line is compatible with the input rating of
the Tester.
The product can receive a nominal power supply voltage in the range of 85 Vac
to 250 Vac at a frequency of 47 Hz or 63 Hz.
The input voltage range is also indicated under LINE VOLTAGE RANGE on
the rear panel.
3.
Connect the power cord to the rear-panel AC inlet, and then connect
the power plug to an outlet that has a ground terminal.
WARNING
NOTE
Summary of Contents for TOS6200A
Page 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Page 14: ...1 8 General TOS6200A 6210...
Page 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Page 34: ...3 8 Part Names and Functions TOS6200A 6210...
Page 76: ...4 42 Basic Operation TOS6200A 6210...
Page 134: ...7 6 Maintenance TOS6200A 6210...
Page 150: ...A 8 Appendix TOS6200A 6210...