Querying Measured Values (SEQuence1:ACQuire)
There are SCALar and ARRay nodes among the second nodes of the MEASure, READ, and FETCh
queries. ARRay is a default node that can be omitted.
You can query the measured value while tests (SEQ2) are being performed.
TRIG:COUN
Sets the trigger count to apply to the SEQuence1 (ACQuire) group.
Command
TRIGger[:SEQuence[1]]:COUNt {<numeric>|MIN|MAX}
TRIGger[:SEQuence[1]]:COUNt?
TRIGger[:ACQuire]:COUNt {<numeric>|MIN|MAX}
TRIGger[:ACQuire]:COUNt?
Parameter
Value:
1 to 100
(The default value is 1.)
Response
Returns the trigger count in <NR3> format.
ABOR
Aborts the tests and measurements of all sequence groups. The trigger state of the TOS5300 Series
immediately after it turns on is the same as its trigger state after it receives an ABOR command.
If you send an ABOR command while the TOS5300 Series is executing tests or measurements, the
measured data is discarded.
If you send an ABOR command without first sending an INIT command and if the measurement
data that is held in the TOS5300 Series is valid, the measured data is not discarded.
Command
ABORt
MEAS:CURR / READ:CURR
Starts a new measurement and queries the current.
Command
MEASure[:ARRay]:CURRent?
READ[:ARRay]:CURRent?
Response
Returns the measured current in <NR3> format or the measured currents as a comma-separated
list in <NR3>,<NR3>,... format.
Unit:
A
Trigger count
Node
Response
1
[:ARRay]
Single measured value
2 or more
[:ARRay]
Comma-separated list of all measured data points
* RST