74
TOS5200
B
Protection Functions
If one or more of the causes of the following problems occur, the protective circuits will acti-
vate, and you will no longer be able to use the TOS5200. This is referred to as PROTECTION
mode.
When the TOS5200 switches to PROTECTION mode, “PROTECTION” on the screen lights.
Follow the instructions below to remove the cause of the problem and clear the PROTEC-
TION mode.
If the TOS5200 has switched to PROTECTION mode because of several causes, the cause
with the highest priority is displayed. After you eliminate the cause, the cause with the next
highest priority is displayed to indicate that there are more causes.
If you cannot clear the mode even by pressing STOP, turn the TOS5200 off and then back on.
If you cannot clear the PROTECTION mode even after you have corrected all the causes, the
TOS5200 may be malfunctioning.
Stop using the product, and contact your Kikusui agent or distributor..
A
kV
M
ȍ
mA
CONFIG
RISE TEST
RMS AVE
MORE CAL RMT
MT
LIMIT
50Hz
60Hz
UPPER
LOWER W COMP
s
1
2
3
READY
TEST
PASS
FAIL
PROTECTION
Display example when an
interlock signal is detected
Panel display
Protection circuit
Description
Remedy
Interlock Protection
An interlock signal was detected.
Remove the interlock signal input,
and then press STOP.
Power Supply Protection An error was detected in the power
supply.
You can clear the PROTECTION
mode by pressing STOP, but if the
protection occurs frequently, repairs
are necessary.
Volt Error Protection
An output voltage outside of the rated
limits was detected.
Withstanding voltage test: ±350 V
Press STOP.
Over Load Protection
An output voltage causing the output
power limit to be executed was
detected.
Withstanding voltage test: 550 VA
Press STOP.
Over Heat Protection
The internal temperature of the
TOS5200 became too high.
Confirm that the internal temperature
of the TOS5200 has decreased, and
then press STOP.
Over Rating Protection
The output current was generated for
a length of time that exceeds the regu-
lated time.
Press STOP.
Remote Protection
A connection to or disconnection from
the REMOTE connector was detected.
Check the REMOTE connector, and
then press STOP.
SIGNAL I/O Protection
The SIGNAL I/O connector’s ENABLE
signal has changed.
Press STOP.
USB Protection
During remote control operation, the
USB connector was disconnected, or
a defect was detected.
Check the USB connector, and then
press STOP.
A
kV
M
ȍ
mA
A
V
M
mA
A
V
M
mA
A
kV
M
ȍ
mA
A
kV
M
ȍ
mA
A
kV
M
ȍ
mA
A
kV
M
ȍ
mA
A
kV
M
ȍ
mA
A
kV
M
ȍ
mA
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......