4
TOS5200
When using this product, be sure to observe the “Safety
Precautions” in the Safety information manual.
When installing this product, be sure to observe the “Pre-
cautions Concerning Installation Location” in the Safety
information manual. The following precautions pertain
only to this product.
• When installing this product, be sure to observe the temper-
ature and humidity ranges indicated below.
Operating temperature range: 0 °C to +40 °C
(32 °F to 104 °F)
Operating humidity range: 20 %rh to 80 %rh
(no condensation)
• When storing this product, be sure to observe the tempera-
ture and humidity ranges indicated below.
Storage temperature range: -20 °C to +70 °C
(-4 °F to 158 °F)
Storage humidity range: 90 %rh or less (no condensation)
• Do not use the product in a poorly ventilated location.
The product uses forced air cooling. It sucks air through the
inlet holes on its right, and left panels, and then expels air
through its rear panel. Secure adequate space around the
product’s inlet and outlet holes to prevent the possibility of
fire caused by accumulation of heat.
Allow at least 20 cm of space between the inlet and outlet
holes on the side and rear panels and the walls (or obsta-
cles).
Hot air (approximately 20°C, 68 °F, hotter than the ambient
temperature) is expelled from the outlet holes. Do not place
objects that are affected by heat near the air outlet.
• Do not use this product near highly sensitive measuring
instruments or receivers.
Noise generated by this product may affect other devices.
At a test voltage of 3 kV or greater, the product may produce
corona discharge between its test lead clips. This will gener-
ate a significant amount of broadband RF emission. To min-
imize this effect, keep the alligator clips away from each
other. Also, keep the alligator clips and test leads away from
conducting surfaces, especially sharp metal edges.
• The TOS5200 Withstanding Voltage Tester is also referred
to as the TOS5200.
• Device under test is also referred to as DUT.
• The term “PC” is used to refer generally to both personal
computers and workstations.
• The following markings are used in the explanations in this
manual.
Indicates a potentially hazardous situation which, if
ignored, could result in death or serious injury.
Indicates a potentially hazardous situation which, if
ignored, may result in damage to the product or other
property.
Indicates information that you should know.
Explanation of terminology or operation principle.
Indicates a reference to detailed information.
SHIFT+key name (blue letters)
Indicates an operation that requires you to press a key
indicated in blue letters while holding down SHIFT.
Indicates useful information.
Safety Precautions
Precautions Concerning Installa-
tion Location
Left panel inlet
holes
Right panel inlet holes
Notations Used in This Manual
WARNING
CAUTION
DESCRIPTION
See
Memo
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......