TOS5200
37
4
Pan
e
l Operation Basics
Panel Memory
You can save up to three sets of test conditions (the test conditions currently being used) to
internal memory.
Test conditions and values that can be saved
Initial panel memory settings
Initially, memory numbers 1 to 3 contain withstanding voltage test conditions. Feel free to use
these test conditions if they match the tests that you want to perform.
For details on the test conditions that are initially contained in memory, see “Default panel
memory values”.
Saving test conditions
1
Display the basic setup screen, and then set the test conditions.
2
Press MEMORY 1, MEMORY 2, or MEMORY 3 to select the memory num-
ber (1 to 3) in which you want to save the current test conditions.
The selected memory number blinks.
3
Press SHIFT+RECALL to save the conditions in the selected memory
number.
After the test conditions have been saved, the basic setup screen is displayed in READY
mode (the READY LED lights).
Test voltage
0.00 kV to 5.50 kV
Output frequency
50 Hz/ 60 Hz
Upper limit
0.01 mA to 110 mA
Lower limit
OFF/ 0.01 mA to 110 mA
Test time
OFF/ 0.1 s to 999 s
Voltage rise time
0.1 s to 10.0 s
Voltage fall time
OFF/ 0.1 s
Start voltage
OFF/ 50 %
A
kV
M
ȍ
mA
CONFIG
RISE
TEST
RMS
AVE
MORE CAL RMT
MT
LIMIT
50Hz
60Hz
UPPER
LOWER W COMP
s
1
2
3
READY
TEST
PASS
FAIL
PROTECTION
See
1
MEMORY
2
3
RECALL
LOCAL
SHIFT
3
2
+
A
kV
M
ȍ
mA
CONFIG
RISE
TEST
RMS
AVE
MORE CAL RMT
MT
LIMIT
50Hz
60Hz
UPPER
LOWER W COMP
s
1
2
3
READY
TEST
PASS
FAIL
PROTECTION
Example of saving to memory
number 1
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......