TOS3200
91
6
Pr
o
g
ram T
est
Storing the Test Results
You can store up to 50 program test results. The storage function is assigned to a function
key while the judgment result (PASS or FAIL) is displayed.
1
Press the Yes (F4) key on the judgment result screen to display the
AUTO TEST DATA screen.
2
Press the arrow keys (
) to select the desired memory number.
3
Press the SAVE (F4) key.
The saved information is displayed.
Press the DATA (F1) key to check the stored data.
4
Press the EXIT (F5) key.
The data save screen closes, and the TOS3200 changes to the READY status.
F5
F4
F3
F2
F1
AUTO 1 / 2
01 PCC––––––P
NRM
•NORM––– 1s
NTWK: B
–––END
PRG
01:TEST-1
TOTAL: 2
PASS
SAVE?
Yes
No
30
μA
0
s
rms
Does not save the test
results.
Skips to step 4.
LINE
0.0W
0.0V 0.00A
F5
F4
F3
F2
F1
F5
F4
F3
F2
F1
DATA SAVE
EXIT
[AUTO TEST DATA]
No
1 2 3 4 5 6 7 8 9 10
00
:
−−UNTITLED−−
2006 / 10 / 17 13 : 45
LIST
10
00:– – UNTITLED– –
2006/10/17 13:45
NETWORK:A MODE:RMS
RANGE:AUTO
FUNC•PRB
POL•COND
CURR TIMER
JUDGE
00 TC•E
NC
P
E
– – P
NRM
•NORM– – 0μA–
1s– PASS
01 TC•E
NC
P
E
– – P
NRM
•NORM– – 0μA–
1s– PASS
– – END
EXIT
Displays the test data of the
selected memory number.
You can check the existing data
before overwriting the existing data.
If you select a memory number in which a test result is already saved, the saved
information is displayed.
Program number: Program name year/month/day hour:minute
Aborts the save operation.
Return to the memory
selection screen.
Move the box using the
arrow keys to select the
number.
Summary of Contents for TOS3200
Page 15: ...General Description This chapter gives an overview and describes the features...
Page 20: ...20 TOS3200 Options This page has been intentionally left blank...
Page 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Page 105: ...Specifications This chapter gives the specifications and external dimensions...
Page 129: ......