PWX
79
Series Operation
Parallel
/Series Operation
4
Settings (series operation)
Setting the voltage and current
Set the voltage and current on each unit. The voltage that is output is the total of the voltages
of the two units. Set the current to the same value on the two units.
Setting the overvoltage protection (OVP) and overcurrent protec-
tion (OCP)
In series operation, you have to set the overvoltage protection (OVP) and overcurrent protec-
tion (OCP) on both units. Specify the same values on the two units.
Starting series operation
Turning the power on and off
Turn the POWER switch on unit 1 and unit 2 on or off.
Turning the output on and off
Press OUTPUT on unit 1 and unit 2 to turn the output on or off.
See
CAUTION
After you turn the POWER switch off, wait at least 10 seconds after the fan stops before you
turn the POWER switch back on. Repeatedly turning the POWER switch on and off at short
intervals can cause damage to the inrush current limiter. In turn, this will shorten the service
life of the POWER switch, internal input fuse, and other components.
Summary of Contents for PWX750ML
Page 6: ...6 PWX Appendix A Options 100 B Troubleshooting 102...
Page 52: ...52 PWX This page is intentionally blank...
Page 80: ...80 PWX This page is intentionally blank...
Page 81: ...Maintenance This chapter explains how to perform cali bration DELETE...
Page 84: ...84 PWX This page is intentionally blank...
Page 99: ...Appendix A Options B Troubleshooting...
Page 106: ...106 PWX This page is intentionally blank...
Page 109: ......