116
PWX
Isolated Analog Interface
Setup and Operation
Using the voltage control interface
1
Use the CONFIG settings to set the following parameters.
2
Connect the external voltage source to the isolated analog interface
connector.
Check that the voltage polarity is correct.
3
Set the applied voltage to the appropriate level.
Using the current control interface
1
Use the CONFIG settings to set the following parameters.
2
Connect the external current source to the isolated analog interface con-
nector.
Check that the current polarity is correct.
3
Set the applied current to the appropriate level.
CAUTION
To prevent damage to the product, do not set the output voltage and output current to values
that exceed the product’s ratings.
See
CONFIG parameter
Setting
CF05
CC control using an external voltage or exter-
nal resistance
on
CF06
CV control using an external voltage or exter-
nal resistance
on
CF09
External control parameter for turning output
on and off
on
CF07 CC or CV control range
Lo (0 V to 5 V) or Hi (0 V to 10 V)
CF08 Range during voltage and current monitoring
Lo (0 V to 5 V) or Hi (0 V to 10 V)
CF10
External control logic parameter for turning
output on and off
Lo (low level signal to turn the output on) or
Hi (high level signal to turn the output on)
1
1.
LOW: 0V to 0.5 V or shorted, HIGH: 4.5 V to 5 V or open.
See
CONFIG parameter
Setting
CF05
CC control using an external voltage or exter-
nal resistance
on
CF06
CV control using an external voltage or exter-
nal resistance
on
CF09
External control parameter for turning output
on and off
on
CF07 CC or CV control range
Hi (0 V to 10 V)
CF08 Range during voltage and current monitoring
Hi (0 V to 10 V)
CF10
External control logic parameter for turning
output on and off
Lo (low level signal to turn the output on) or
Hi (high level signal to turn the output on)
1
1.
LOW: 0V to 0.5 V or shorted, HIGH: 4.5 V to 5 V or open.
Summary of Contents for PWX1500H
Page 10: ...10 PWX This page is intentionally blank...
Page 26: ...26 PWX This page is intentionally blank...
Page 88: ...88 PWX This page is intentionally blank...
Page 89: ...Maintenance This chapter explains how to perform cali bration DELETE...
Page 92: ...92 PWX This page is intentionally blank...
Page 120: ...120 PWX This page is intentionally blank...
Page 121: ...Appendix A Options B Troubleshooting...