MEASure/FETCh Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
INST:INFO
Queries the information of the channel currently being controlled.
The channel under control is set with the
Command
INSTrument:INFO?
Response
Returns the maximum voltage, maximum current, maximum power, and model
name of the channel under control in response to INST:INFO?.
(Example) If the channel under control is a PWR401ML (maximum voltage: 80 V,
maximum current: 20 A, maximum power: 400 W)
+8.0000E+01, +2.0000E+01, +4.0000E+02, PWR401ML
is returned.
MEASure/FETCh Command
This product does not have the INITiate or FETCh measurement function.
The measured voltage and the measured current are updated alternately at 25 ms
intervals.
MEASure and FETCh are aliases and operate in the same manner. Queries the lat-
est measured value when the command is sent.
-> Tutorial “Settings and Measurement”
FETC:ALL/ MEAS:ALL
Queries the current and voltage.
When you are using the multichannel function, use <n> to specify the channel num-
ber.
Command
FETCh[<n>][:SCALar]:ALL?
MEASure[<n>][:SCALar]:ALL?
(Example)
MEAS:ALL?
MEAS2:ALL?
Response
Returns the current <NR3> and then the voltage <NR3> in a comma-separated
format in response to FETC:ALL?/ MEAS:ALL?.
When multichannel is in use, this command returns the current <NR3> and then
the voltage <NR3> of the specified channel in a comma-separated format in re-
sponse to FETC<n>:ALL?/ MEAS<n>:ALL?. If you omit <n>, the measured value
of the channel that was specified by the INST command is returned. The measured
voltage and the measured current are updated alternately at 25 ms intervals.
Unit:
Current
A
Voltage
V