User’s Manual
PMX_MULTI
Specifications
Other functions
Common specifications
Output-on delay/ off delay
Applicable channels
All channels
Setup
Set the output on/off delay time.
Setting range
0.1 s to 99.9 s
Resolution
0.1 s
Setting accuracy
*1
*1.
The difference between the time from when the reference output reaches 50 % of the setting to
when the target output reaches 50 % of the setting and the delay time setting.
±50 ms
Memory
Saves three combinations of voltage, current, OVP, OCP,
and output-on delay/ off delay settings.
key lock
Selectable from the following three modes.
• Loc1: Lock all keys except the OUTPUT and memory A,
B, and C keys.
• Loc2: Lock all keys except the OUTPUT key.
• Loc3: Lock all keys and the rotary knob.
Tracking
Applicable channels
All channels
Operation
mode
Tracking function 1
*2
*2.
In tracking function 1, the output can be varied within the output range of the reference channel
voltage or current.
Absolute value change
Tracking function 2
*3
*3.
In tracking function 2, the output can be varied at the same percentage as the reference output in
reference to the output at the start of the tracking function.
Percentage change
Setting
accuracy
CV setting accuracy
0.4 % of 40 mV
CC setting accuracy 0.7 % of 10 mA