82
User’s Manual
PMX_MULTI
Maintenance
Connection (series operation and parallel operation)
Connection for voltage calibration (for series operation)
1
Turn the POWER switch off.
2
Connect a DVM to the output terminals.
Connection for current calibration (for parallel operation)
1
Turn the POWER switch off.
2
Connect a shunt or ammeter to the out-
put terminals, and connect a DVM
across it.
WARNING
Risk of electric shock.
• Turn the POWER switch off before you touch the output terminals.
• Connect the output terminal to the chassis terminal.
PMX
Output terminal
CH2
Output terminal
CH1
Chassis terminal
Connect the negative terminal
of CH1 or the positive terminal
of CH2 to the chassis terminal.
DVM
HI
LO
HI
LO
PMX
Output terminal
CH2
Chassis terminal
Output terminal
CH1
Chassis terminal
Shunt or ammeter
DVM
Connect the output terminal
of one of the channels to the
chassis terminal.