PMX_MULTI
User’s Manual
81
Maintenance
Output current calibration (offset and full scale)
1
Press SELECT to select the channel you want to calibrate.
2
Use the VOLTAGE knob to select
SC-B
, and turn on the output.
3
Turn the CURRENT knob to set the DVM reading (converted current
value) to 1 % of the rated output current.
4
Use the VOLTAGE knob to select
SC-T
, and turn on the output.
5
Turn the CURRENT knob to set the DVM reading (converted current
value) to 100 % of the rated output current.
Ammeter calibration (offset, full scale, and OCP)
Be sure to calibrate the output current before calibrating the ammeter.
1
Press SELECT to select the channel you want to calibrate.
2
Use the VOLTAGE knob to select
DC-B
, and turn on the output.
3
Turn the CURRENT knob to match the product’s current reading to the
DVM reading.
4
Use the VOLTAGE knob to select
DC-T
, and turn on the output.
5
Turn the CURRENT knob to match the product’s current reading to the
DVM reading (converted current value).
6
Turn the VOLTAGE knob to select
OCPC
.
7
Press OVP•OCP.
Calibration is performed automatically. The voltmeter and ammeter displays change
during this process.
8
If the calibration completes successfully, the ammeter displays
“
DONE
.”
If it doesn’t, the ammeter displays “
FAIL
.”
If the calibration ends with any other display, contact your Kikusui agent or distribu-
tor.
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
If you press OUTPUT ALL during the calibration, the OCP calibration ends, and the ammeter
displays “
FAIL
.”