80
User’s Manual
PMX_MULTI
Maintenance
Output voltage calibration (offset and full scale)
1
Press SELECT to select the channel you want to calibrate.
2
Use the VOLTAGE knob to select
SV-B
, and turn on the output.
3
Turn the CURRENT knob to set the DVM reading to 1 % of the rated
output voltage.
4
Use the VOLTAGE knob to select
SV-T
, and turn on the output.
5
Turn the CURRENT knob to set the DVM reading to 100 % of the rated
output voltage.
Voltmeter calibration (offset, full scale, and OVP)
Be sure to calibrate the output voltage before calibrating the voltmeter.
1
Press SELECT to select the channel you want to calibrate.
2
Use the VOLTAGE knob to select
DV-B
, and turn on the output.
3
Turn the CURRENT knob to match the product’s voltage reading to the
DVM reading.
4
Use the VOLTAGE knob to select
DV-T
, and turn on the output.
5
Turn the CURRENT knob to match the product’s voltage reading to the
DVM reading.
6
Turn the VOLTAGE knob to select
OVPC
.
7
Press OVP•OCP.
Calibration is performed automatically. The voltmeter and ammeter displays change
during this process.
8
If the calibration completes successfully, the ammeter displays
“
DONE
.”
If it doesn’t, the ammeter displays “
FAIL
.”
If the calibration ends with any other display, contact your Kikusui agent or distribu-
tor.
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
CV
V
s
OFF
CC
A
s
ON
PARA
SERI
TRACKING 12
%
%
If you press OUTPUT ALL during the calibration, the OVP calibration ends, and the ammeter
displays “
FAIL
.”