46
User’s Manual
PMX_MULTI
Basic Functions | CONFIG Settings
CF03 (
LOC
)
Key lock content
The key lock function locks the panel. Select from the following three values.
CF04 (
E.OUT
)
External control for turning output on and off
Enables or disables output on/off control using an external contact
.
Can be set separately for each channel.
CF05 (
E.LOG
)
External control logic for turning output on and off
Sets the logic that is used for output on/off control using an external contact
Can be set separately for each channel.
CF06 (
PRIO
)
Output-on startup state
Sets the prioritized operation mode when the output is turned on
. Can be set separately for each
channel.
CF07 (
RIS.T
)
Rise time when a value is changed
You can set the rise time that is applied when you change a value
. Can be set separately for each
channel.
If you specify “
FAST
,” the rising waveform may become unstable and distort.
Display
Description
LOC1
Lock1
Only the OUTPUT key and recalling of memory (A/B/C) are
allowed.
LOC2
Lock2
Only the OUTPUT key can be used.
LOC3
Lock3
All keys and rotary operations are disabled (factory default set-
ting).
Display
Description
OFF
OFF
Disabled (factory default setting)
ON
ON
Enabled
Display
Description
LO
LOW
The output is turned on with a LOW signal (0 V to 0.5 V) or short
circuit (factory default setting).
HI
HIGH
The output is turned on with a HIGH signal (4.5 V to 5 V) or open
circuit.
Display
Description
CV
CV
CV (constant voltage) is prioritized (factory default setting).
CC
CC
CC (constant current) is prioritized.
Display
Description
SLOW
SLOW
The value rises in 100 ms (factory default setting).
FAST
FAST
The value rises in 10 ms.