PLZ-4WL
33
Constant Resistance Mode (CR)
2
Basi
c Funct
ions
6
Turn the rotary knob to set the conductance.
The CR mode settings are complete. When you press LOAD, the LOAD LED lights, and the
PLZ-4WL starts supplying current. You can change the conductance even when the load is
on.
Configuring CR+CV mode
CR+CV mode is constant resistance mode (CR mode) with CV mode added to it. You can
add CV mode even while the load is on.
1
Configure the constant resistance settings.
To configure the settings, refer to "Configuring CR mode".
2
Press +CV (SHIFT+MODE) to add CV mode.
The PLZ-4WL switches to CR+CV mode, and “CR+CV” appears on the display.
3
Make sure that the PLZ-4WL is in the fundamental setting entry
condition.
If SET is not highlighted, press SET/VSET to switch to the fundamental setting entry
condition.
4
Turn the rotary knob to set the conductance and the voltage.
The value that you can set (conductance or voltage) changes when you press SET/
VSET.
The CR+CV mode settings are complete. Press LOAD to start testing. You can change the
conductance and voltage even when the load is on.