PFX2731S
Operation Manual
41
Specifications
Electric Specifications
Unless specified otherwise, the specifications are for the following settings and conditions.
• The product is warmed up for at least 30 minutes.
The used terminology is as follows:
• TYP: These are typical values that are representative of situations where the product operates in an environment
with an ambient temperature of 23 °C (73.4 °F). These values do not guarantee the performance of this product.
• setting: Indicates a set value.
• reading: Indicates a readout value.
• rating: Indicates a rated value.
• Static: General term to indicate CC charge, CC - CV charge, CC discharge, CC - CV discharge, CP discharge, and
CP - CV discharge.
• Pattern: General term to indicate pattern charge/discharge.
C-rate entry is supported.
• Individual CC values, pattern charge/discharge step current, limit current, current capacity of end condition/It cutoff
current can be set using a C-rate.
• C-rate calculation specifications
The nominal capacity setting value is assumed to be 1C.
C-rate setting range: 0.001 to 99.999 (the range is common.)
The conversion result from a C-rate to a current/capacity value is rounded to the nearest control resolution.
Rated output
DUT cable: 5.5 mm
2
in diameter, 5 min length
Item
Specifications
Number of outputs
6 ch (per frame)
Isolation
Between channels, Ch-chassis
Charge current range
2 A range
0.0000 A to 2.0000 A
20 A range
0.000 A to 20.000 A
Charge voltage range
-1.000 V to 6.000 V
Discharge current
range
2 A range
0.0000 A to 2.0000 A
20 A range
0.000 A to 20.000 A
Discharge voltage range
-1.000 V to 6.000 V