PFX2512
43
2
C
o
nnec
ting and C
o
nfiguring De
vic
e
s
Connecting the DUT (Battery)
We recommend that you install a fuse between the PFX2512 and DUT (battery) to allow safe
connection between them and to prevent short-circuit accidents in DUT cables and voltage
sensing cables. Insert the fuse as close to the DUT (battery) terminal section as possible.
WARNING
Risk of electric shock.
• After connecting the devices, connect the PFX2512 to the DUT (battery). Connect the
PFX2512 side first.
• When installing a fuse between the PFX2512 and DUT (battery), be sure to turn off the
POWER switch, and turn off the circuit breaker of distribution.
CAUTION
Be sure to connect the voltage sensing cable.
The system will not work properly if it is not connected. Moreover, make sure to connect the cable
close to the DUT (battery). Performing a charge/discharge test without connecting the voltage
sensing cable may damage the DUT (battery) and is extremely dangerous.
Connect the DUT (battery) after turning on the power switch.
If there is a problem, the warning function (
) can prevent failure from occurring.
See
DUT cable
Voltage sensing cable
Temperature sensing cable
DUT (battery)
Thermistor
+
–
Install the fuse as close
to the DUT (battery)
terminal as possible.
Fuse for the DUT cable
120 % of the maximum
charge/discharge current
DUT (battery)
Fuse for the voltage sensing cable
less than 0.1 A
Input/Output terminal board
PFX2512
Example for the installation of fuse between the PFX2512 and the DUT (battery)
+
+
S
+
DUT
–
–
DUT
–
S
+
T
–
T
• DUT cable fuse rating:
120 % of the maximum charge/discharge current
• Voltage sensing cable fuse rating:
0.1 A or less
CAUTION
• A fuse blowing may lead to explosions or fire. Be sure to put a cover on the fuse.
• Use DC fuses that have sufficient breaking capacity.
Summary of Contents for PFX2500 Series
Page 12: ...12 PFX2512 This page is intentionally blank ...
Page 62: ...62 PFX2512 This page is intentionally blank ...
Page 71: ...Specifications This chapter provides the PFX2512 s specifications and outline drawing ...
Page 96: ...96 PFX2512 This page is intentionally blank ...
Page 140: ...140 PFX2512 Troubleshooting This page is intentionally blank ...
Page 143: ......