Cleaning the Dust Filter
A dust filter is installed on the inside of the louver. Periodically clean the filter to prevent clogging.
1
While pulling the bottom slat towards you,
slide the top of the louver down, and remove
the louver from the panel.
2
Remove the dust filter from the inside of the
louver and clean it.
Use a vacuum cleaner to dispose of the dust and
foreign particles that are attached to the dust filter.
If the filter is extremely dirty, clean it using water-
diluted neutral detergent, and dry it completely.
3
Attach the dust filter to the louver.
Attach the dust filter so that the louver’s tabs pass
through the cuts in the dust filter.
4
Insert the dust filter into the louver, and then
attach the louver so that its tabs enter the slots
in the panel. While pushing on the second slat
from the bottom of the louver, slide the louver
up to attach it.
The pin will lock in place.
WARNING
Risk of electric shock. When performing maintenance work, be sure to turn off the POWER
switch and remove the power cord plug.
CAUTION
Clogging may lead to malfunction or the reduction of the product’s service life.
• If the dust filter is clogged, the product’s internal cooling capabilities will be reduced.
• When the PFX2512 is in operation, air is sucked through the dust filter to cool the inside of the
device. If moisture is present in the dust filter, the temperature or humidity inside the PFX2512
increases.
Dust filter
Louver
Tab
Pin
Cut
Fix the pin inside the
louver in place.
Summary of Contents for PFX2500 Series
Page 12: ...12 PFX2512 This page is intentionally blank ...
Page 62: ...62 PFX2512 This page is intentionally blank ...
Page 71: ...Specifications This chapter provides the PFX2512 s specifications and outline drawing ...
Page 96: ...96 PFX2512 This page is intentionally blank ...
Page 140: ...140 PFX2512 Troubleshooting This page is intentionally blank ...
Page 143: ......