PCZ1000A
3-23
3
Ope
ration
3.7.2
Parallel Operation
The front panel display on the master unit shows the total current for the units
connected in parallel.
The measurement accuracy can be improved by performing calibration in the
parallel configuration.
The setting resolution for parallel operation is different depending on the number of
instruments used.
Ammeter accuracy (RMS) : ± 2 % of FS
Ammeter accuracy (PEAK): ± 4 % of FS
page 3-24
Always keep the load wiring away as possible from the flat cable to prevent possible
unstable operation.
Fig.3-8 Parallel Connection with Two Slave Units
Table 3-2 shows the relation between the capacity and the number of parallel units.
Table 3-2
Capacity and Number of Parallel Units
See
PARALLEL IN
PARALLEL IN
PARALLEL OUT
PARALLEL OUT
L
L
L
N
N
N
G
G
G
G
L
Equipment
under test
Flat cable
Master
Slave
Slave
N
No. of Slave Units
Maximum Current
(RMS)
Maximum Current
(Peak)
Maximum Power
1
20 A
80 A
2 000 W
2
30 A
120 A
3 000 W
3
40 A
160 A
4 000 W
4
50 A
200 A
5 000 W
Summary of Contents for PCZ1000A
Page 4: ...ii PCZ1000A This page is intentionally blank ...
Page 17: ...General Description This chapter gives an overview and describes the features of the PCZ1000A ...
Page 35: ...Operation This chapter describes each mode and function of the operation ...
Page 82: ...5 4 PCZ1000A This page is intentionally blank ...
Page 83: ...Specifications This chapter gives the specifications and dimensions of the PCZ1000A ...
Page 88: ...6 6 PCZ1000A This page is intentionally blank ...
Page 89: ...The appendix contains the basic operating modes operating range and troubleshooting Appendix ...
Page 99: ......