KIKUSUI Electronics Corp.
96
FETC:ARR:<harm-item>?/ MEAS:ARR:<harm-item>?
Queries the harmonic data specified with <harm-item>.
Command
FETCh:ARRay:<harm-item>? [(@chanlist)]
MEASure:ARRay:<harm-item>? [(@chanlist)]
Measurement <harm-item> list of each phase
Phase designation (@chanlist) possible
0th to 50th
<harm-item>
Description
Unit
CURRent:HARMonic[:AMPLitude]
Harmonic current
Arms
CURRent:HARMonic:PHASe
Harmonic current phase
Deg
CURRent:HARMonic:THD
Total current harmonic distortion PCT
VOLTage:HARMonic[:AMPLitude]
Harmonic voltage
Vrms
VOLTage:HARMonic:PHASe
Harmonic voltage phase
Deg
VOLTage:HARMonic:THD
Total voltage harmonic distortion PCT
Response
Returns the measurements (0th to 50th) of the designated phase in a com-
ma-separated NR3 format in response to FETC:ARR:<harm-item>? [(@chanlist)]/
MEAS:ARR:<harm-item>? [(@chanlist)].
Returns the measurements (0th to 50th) of the phase designated by the INST
command in a comma-separated NR3 format in response to FETC:<meas-item>?/
MEAS:<meas-item>?.
Returns the measurements (0th to 50th) of the U phase in a comma-separated
NR3 format when multiple phases are designated. Then, the V-phase measure-
ments and W-phase measurements are returned in the same manner.
Example
INST:NSEL 1;COUP NONE
FETC:ARR:CURR:HARM?
Returns the harmonic current of the U phase.
FETC:ARR:CURR:HARM? (@2)
Returns the harmonic current of the V phase.
FETC:ARR:CURR:HARM? (@1:3)
Returns the harmonic currents of all phases.