KIKUSUI Electronics Corp.
238
STAT:QUES:INST:ISUM{1|2|3}
Queries the event of the QUEStionable:INSTrument:ISUMmary{1|2|3} subregister.
This command is valid for single-phase three-wire output and three-phase output.
A query clears the contents of the register.
Of the parameters {1|2|3}, 1 represents U phase, 2 represents V phase, and 3 rep-
resents W phase.
Command
U phase
STATus:QUEStionable:INSTrument:ISUMmary1[:EVENt]?
V phase
STATus:QUEStionable:INSTrument:ISUMmary2[:EVENt]?
W phase (three-phase output only)
STATus:QUEStionable:INSTrument:ISUMmary3[:EVENt]?
Response: NR1