[SOURce:]VOLTage Command
PCR-WE Interface Manual
171
[SOURce:]VOLTage Command
VOLT
Set the AC voltage.
This command is invalid when the combined value with the DC voltage is outside the
allowable range (L range: -222.5 V to 222.5 V, H range: -445.0 V to 445.0 V).
If the signal source is set to external signal source (VOLT:EXT:EXTDC:SIGN:SOUR
EXT), this command is invalid.
This command is invalid when an external analog signal is used to control the voltage
or frequency (VOLT:EXT:FUNC:MODE VPR).
Command
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <numeric>[,(@chan-
list)]
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [(@chanlist)]
Parameter
Value:
0 to 157.5 AC voltage in the L range
0 to 315.0 AC voltage in the H range
(The default value is 0.)
Unit:
V
Settings are reset to default values when an *RST command is sent.
Example
VOLT 120
Response: NR3