PROGram Command
PCR-WE Interface Manual
129
PROG:EDIT:PHAS:UV
Sets the U-V phase difference of the sequence step.
Use PROG:EDIT:PHAS:RAMP to set the phase signal change.
This command is valid for three-phase output.
Command
PROGram:EDIT:PHASe:UV <step_NRf>,<enable_boolean>,<phase_NR3>
PROGram:EDIT:PHASe:UV? <step_NRf>
Parameter
<step_NRf>
Value: 0 to 599 Number of the step that you want to configure
Parameter
<enable_boolean>
Value:
ON(1)
Enables U-V phase difference control
OFF(0) Disables U-V phase difference control (default)
Parameter
<phase_NR3>
Value:
0.00 to 360.00
U-V phase difference (The default value is 0)
0 and 360 are the same.
Unit:
DEG
Example
PROG:EDIT:PHAS:UV 1,ON,90
PROG:EDIT:PHAS:UV? 1
Response
Returns the U-V phase difference of the specified step in the following order.
U-V phase difference control enabled/disabled<NR1>, U-V phase difference<NR3>