PAT-T 4KW
7-23
7
Remot
e
Cont
ro
l
Response
Returns the current setting in the NR3 form in response to CURR?. If the current
value is set using external input, the specified current is returned.
7.9
Measurement Operation Settings
For the procedure to set the measurement operation, See “Measurement operation”
on page 7-39.
7.10 Protection Function Settings
Overvoltage protection settings
VOLT:PROT
Sets the overvoltage protection (OVP).
Command
[SOURce:]VOLTage:PROTection[:LEVel]
{<numeric>|MIN|MAX}
[SOURce:]VOLTage:PROTection[:LEVel]? [MIN|MAX]
Parameter
Val
u
e:
10 % to 111.5 % of the rated output voltage
(The default value is 111.5 % of the rated output voltage.)
Unit
V
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the OVP value in the NR3 form in response to VOLT:PROT?.
Voltage limit setting
VOLT:LIM
Queries the voltage limit setting. The voltage limit setting is approximately 95 % of
the OVP trip point.
This command is valid when VOLT:LIM:AUTO ON is specified.
Command
[SOURce:]VOLTage[:AMPLitude]?
Response
Returns the voltage limit setting value in the NR3 form in response to VOLT:LIM?.
Unit
V
Summary of Contents for PAT160-25T
Page 17: ...General Description This chapter describes an overview and describes the features...
Page 22: ...1 6 PAT T 4kW...
Page 34: ...2 12 PAT T 4kW...
Page 106: ...6 12 PAT T 4kW...
Page 158: ...7 52 PAT T 4KW...
Page 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Page 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...