PAN-A
4-1
Chap
.4
Applied Oper
ation
1
Chapter 4
Applied Operation
Use of the CONTROL terminal board on the rear panel allows the
unit to perform remote sensing, external output control, parallel oper-
ation, and series operation.
• To avoid electric shock, for remote sensing lines and
control lines, use wires with a rated voltage higher
than the isolation voltage of the unit. If you cannot
obtain such wires, protect the wires by using insula-
tion tubes with a withstand voltage higher than the
isolation voltage of the unit.
4.1
CONTROL terminal board
This section describes the procedure for connecting wires, and pre-
cautions when using the CONTROL terminal board.
■
Wires and tools required for connection
1.Wires
As shown in Fig.4-1, the CONTROL terminal board has two
types of terminals, which use different wires.
Group A (No. 2 to 19 terminals)
Single wire:
φ
0.32 to
φ
0.65 (AWG28 to AWG22)
Twisted wire: 0.08 mm
2
to 0.32 mm
2
(AWG28 to AWG22)
Group B (-S, +S, and No. 1 terminal)
Single wire:
φ
0.4 to
φ
1.2 (AWG26 to AWG16)
Twisted wire: 0.3 mm
2
to 1.25 mm
2
(AWG22 to AWG16)
(Bare wire diameter:
φ
0.18 or more)
2.Screwdriver
Axis diameter:
φ
3
End width: 2.6 mm
3. Wire stripper
One suitable for the wires described above
WARNING