PAD-L
Basic Operation
3-11
3.4
Using the guard cap(s)
The unit includes VOLTAGE and CURRENT setting knobs; in order to make them
mechanically fixed or semi-fixed, guard caps are supplied with the unit. Use them if
output presetting is seldom necessary.
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
NOTE
•
For the following models, the CURRENT setting knob is of the double-piece
type; the guard caps cannot be used.
PAD16-100L
All TYPE IV models
All TYPE V models
•
If you intend to reinstall the knob, keep it in a safe place.
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
Mounting procedure
1.
Turn on the POWER switch.
2.
While pressing the CURRENT/VOLT. LIMIT switch, set the output level
to a desired value. Here, a rough presetting is acceptable.
3.
Use the supplied allen wrench to remove an output setting knob that is to
be replaced with a guard cap.
4.
While pressing the CURRENT/VOLT. LIMIT switch, set the output level
to a desired value by using a slotted-head screwdriver.
•
To fix presetting
5.
Mount the guard cap, instead of the removed knob.
6.
Press the CURRENT/VOLT. LIMIT switch again to confirm the
presetting is not changed.
•
To semi-fix presetting
5.
As shown in Fig. 3-4, pierce the guard cap top with a phillips-head
screwdriver.
6.
Mount the guard cap, instead of the removed knob.
Since the guard cap is pierced, the output level may be readjusted using a
slotted-head screwdriver.