3
KHA3000
67
W
iring M
easurem
ent Circuits
Measurement Circuit Using Commercial
Power Supply
Connect this product to the commercial power supply directly. In directly connecting the
SOURCE terminal to the commercial power supply (AC line) without an AC power supply,
such as when measuring in-rush current, use a circuit breaker that can disconnect this
product from the commercial power supply (AC line) for safety (diagram shown below).
Parts to be prepared by the customer
•
Circuit breaker
•
Connection cable with commercial power supply
See
WARNING
•
You may receive an electric shock. To install a circuit breaker between the SOURCE
terminal and commercial power supply (AC line), be sure to shut off the power
supply from the switchboard by turning off the switches on the switchboard.
•
Use the circuit breaker with the rated current set to 40 Arms (maximum input current
of this product) or less.
•
Use a circuit integrating three circuit breaker poles so that the U, V and W poles can
be shut down simultaneously.
A
V
A
V
A
V
Commercial power supply
N
G
U
V
W
N
G
U
V
W
N
G
L1
L2
L3
SOURCE
LOAD
Circuit breaker
KHA3000
EUT
CAUTION
Terminal G of this product is not a protective conductor terminal. It is not grounded even if
the wire is connected to terminal G. For terminals G and N, the respective SOURCE and
LOAD terminals are short-circuited internally.
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......