Common Information on Measurement Circuits
Prevent an Electric Shock with Terminal Cover
To prevent an electric shock, covers are attached to the LOAD and SOURCE terminals. Use
these terminals as shown in the figure below.
Performance of Input Terminals
●
Maximum input voltage
The maximum input voltage at the voltage input
terminal and current input terminal is 600 Vms/
900 Vpeak. This product cannot be used if the
maximum voltage is exceeded.
●
Maximum input current
The maximum input current at the current input
terminal is 40 Arms/100 Apeak. This product
cannot be used if the maximum current is
exceeded.
●
Instrument loss (input resistance)
The instrument loss (input resistance) of the
voltage input terminal is about 4.7 M
, and that
of the current input terminal is about 3.5 m
.
●
Measurement category
The voltage input terminal and current input
terminal conform to IEC standard measurement
category CAT I or CAT II. The measurement
category varies depending on the input voltage.
WARNING
•
To prevent an electric shock, attach the terminal cover.
Spring washer (M6)
Knob
Terminal cover
Clamping material
M6
Attach the terminal cover so that
a wire can be placed in this part.
LOAD
N/L2
L1
V
+
-
L1
40A rms
MAX
EXT INPUT
EXT CLAMP
SOURCE
VOLTAGE INPUT
CURRENT
INPUT
VOLTAGE
SENSING / INPUT
5Vpk MAX
+
-
Current input
terminals
Voltage input
terminals
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......