358
KHA3000
Specifications
Harmonic Current Measurement Function
Item
Specifications
Conforming standard
IEC 61000-3-2 Ed5.0, IEC 61000-3-2 Ed3.0, IEC 61000-3-2Ed2.2,
JIS C61000-3-2(2011), JIS C61000-3-2(2005)
IEC 61000-3-12 Ed2.0, IEC 61000-3-12 Ed1.0
Requirements for measuring
instrument
*1
*1.
Applicable when the current range is 1 A to 40 A and fundamental current input is within the effective input
range. The current that satisfies IEC 61000-4-7 Ed2.0 or IEC 61000-4-7 Ed2.1 is 37.5 Arms or less (single-
phase) or 40 Arms/phase or less (three-phase).
IEC 61000-4-7 Ed2.1(2009), IEC 61000-4-7 Ed2.0(2002),
IEC 61000-4-7 Ed1.0(1991)
Harmonic
analysis order
HA mode
40th order
OTHER mode
180th order
*2
*2.
When the FFT analyzer function is used.
Accuracy
*3
*3.
The minimum harmonic current value is 3 % or higher of the current range and fundamental current input
is within the effective input range. Specified for the ambient temperature of 18
°C
to 28
°C
and relative
humidity of 20 % to 80 % (without condensation). When scaling is used, the voltage range shown on the
display is the "range" value.
45 Hz to
65 Hz
0.5 A
range
±
(0.5 % of rdng+0.2 % of range)
1 A to 40 A
range
±
(0.5 % of rdng+0.1 % of range)
66 Hz to
2.4 kHz
0.5 A
range
±
((0.5+0.417
×
n kHz)% of rdng+0.2 % of range)
: n is frequency.
1 A to 40 A
range
±
((0.5+0.417
×
n kHz)% of rdng+0.1 % of range)
: n is frequency.
2.4 kHz to 9 kHz
Accuracy not specified.
Resolution
0.5 A to 40 A range
Same as current measurement function
Interharmonics processing
Processing ON (2nd order and more): IEC 61000-4-7 Ed2.1(2009)
Processing ON: IEC 61000-4-7 Ed2.0(2002)
Processing OFF: IEC 61000-4-7 Ed1.0(1991)
Window function
Rectangular
Window width 10-wave (50 Hz)
IEC 61000-4-7 Ed2.1(2009), IEC 61000-4-7 Ed2.0(2002)
12-wave (60 Hz)
16-wave
(50 Hz/60 Hz)
IEC 61000-4-7 Ed1.0(1991)
Number of sampling points
Fixed to 16384
Sampling rate
*4
*4.
Varies in synchronization with the AC power frequency.
Up to 106.5 ksps
FFT processing word length
32 bit
Anti-aliasing
filter
HA mode
Cutoff frequency = 6 kHz 4th Butterworth type
OTHER mode
Cutoff frequency = 15 kHz 4th Butterworth type
Stop Band
Attenuation
The attenuation defined at the fundamental frequency of 50 Hz/60
Hz, against for the aliasing in the range from the fundamental to the
40th harmonic is more than -50 dB.
1.5 second filter
Time constant 1.5 second digital low pass filter processing for the
effective values of voltage and current, real power and harmonic
current
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......