356
KHA3000
Specifications
Current Measurement Function
External Current Input Terminal (EXT INPUT)
Item
Specifications
Range
0.5 A/1 A/2 A/5 A/10 A/20 A/ 40 A
Allowable crest
factor
0.5 A to 20 A range 4
40 A range
2.5 and 4 (up to 20 ms)
Display item
TrueRMS and ± peak
Accuracy
*1
*1.
Accuracy is not specified for
±
Peak. PLL is in the Lock state, in the effective input range, and with anti-
aliasing filter cutoff frequency 6 kHz. Specified for the ambient temperature of 18
°C
to 28
°C
and relative
humidity of 20 % to 80 % (without condensation). When scaling is used, the voltage range shown on the
display is the "range" value.
45 Hz to
65 Hz
0.5 A
range
±
(0.5 % of rdng+0.2 % of range)
1 A to
40 A
range
±
(0.5 % of rdng+0.1 % of range)
66 Hz to
2.4 kHz
0.5 A
range
±
((0.5+0.417
×
n kHz)% of rdng+0.2 % of range)
: n is frequency.
1 A to
40 A
range
±
((0.5+0.417
×
n kHz)% of rdng+0.1 % of range)
: n is frequency.
2.4 kHz to 9 kHz
Accuracy not specified.
Resolution
(RMS/Peak)
0.5 A range
0.0001 A/0.0001 A
1 A range
0.0001 A/0.0002 A
2 A range
0.0001 A/0.0004 A
5 A range
0.0001 A/0.001 A
10 A range
0.001 A/0.002 A
20 A range
0.001 A/0.004 A
40 A range
0.001 A/0.008 A
Display range
(RMS/Peak)
0.5 A to 10 A range
120 % of the range/400 % of the range
Effective input
range
0.5 A range
50 % to 110 % of the range
1 A to 40 A range
20 % to 110 % of the range
Excess input display
“
Over
”
is displayed when the
±
peak value or RMS value exceeds
the display range.
Item
Specifications
Input method
Voltage input (BNC terminal)
Input impedance
1 M
±
10 %
Maximum input voltage
±
5 Vpeak
Display current/input voltage ratio
±
20 Apeak/1 Vpp (40 Arms/1 Vrms)
Display current/
input voltage ratio
(for each range)
*1
*1.
Accuracy is not specified for
±
Peak. PLL is in the Lock state, in the effective input range, and with anti-
aliasing filter cutoff frequency 6 kHz.
0.5 A range
±
2.0000 Apeak/0.1000 Vpp
1 A range
±
4.0000 Apeak/0.2000 Vpp
2 A range
±
8.0000 Apeak/0.4000 Vpp
5 A range
±
7.0710 Apeak/0.3535 Vpp
(± 20.0000 Apeak/1.0000 Vpp or equivalent)
10 A range
±
40.0000 Apeak/2.000 Vpp
20 A range
±
80.0000 Apeak/4.000 Vpp
40 A range
±
160.0000 Apeak/8.000 Vpp
Measurement accuracy
*2
*2.
Specified for the ambient temperature of 18
°C
to 28
°C
and relative humidity of 20 % to 80 % (without
condensation).
± (
| Current measurement accuracy | + External shunt ac
0.5 % of range)
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......