Voltage Fluctuation Measurement Mode (61000-3-3)
Rem
ote Contr
ol
Setting the Margins
VFL:IEC33:MARG:D
Sets the margin for the limit value of dc, dmax or Tmax (d(t)>3.3%). Set a relative value of the
reference limit value (100). The setting range is 10 to 100. For instance, specify 80 to set the
margin to 80 % of the standard limit value. For Tmax (d(t)>3.3%), set a margin for the limit
value of the time exceeding Tmax (d(t)>3.3%).
Command
VFLuctuation:IEC33:MARGin:D {<numeric>|MIN|MAX}
VFLuctuation:IEC33:MARGin:D?
Parameter Value:
10 to 100 (The default is 100.)
Unit:
PCT
Response
Returns the margin for the limit value of dc, dmax, or Tmax (d(t)>3.3%) in the <NR3> format.
VFL:IEC33:MARG:FLIC
Sets the margin for the limit value of Pst or Plt. Set a relative value of the standard limit value
(100). The setting range is 10 to 100. For instance, specify 80 to set the margin to 80 % of the
standard limit value.
Command
VFLuctuation:IEC33:MARGin:FLICker {<numeric>|MIN|MAX}
VFLuctuation:IEC33:MARGin:FLICker?
Parameter Value:
10 to 100 (The default is 100.)
Unit:
PCT
Response
Returns the margin for the limit value of Pst or Plt in the <NR3> format.
** RST
RST
** RST
RST
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......