Voltage Fluctuation Measurement Mode (61000-
3-3)
Measurement Time, Measurement Count and Limit Values
VFL:IEC33:DMAX
Sets the limit value of dmax.
Command
VFLuctuation:IEC33:DMAXlimit {4|6|7}
VFLuctuation:IEC33:DMAXlimit?
Parameter Value:
4, 6 or 7 (The default is 6.)
Unit:
PCT
Response
Returns the limit value of dmax (maximum relative voltage changes) in the <NR3> format.
VFL:IEC33:DCO
Sets the d measurement count.
Command
VFLuctuation:IEC33:DCOunt <NRf>
VFLuctuation:IEC33:DCOunt?
Parameter Value:
3 to 24 (The default is 24.)
Unit:
None
Response
Returns the d measurement count in the <NR3> format.
VFL:IEC33:DTIM
Sets the d measurement time.
Command
VFLuctuation:IEC33:DTIMe {<numeric>|MIN|MAX}
VFLuctuation:IEC33:DTIMe?
Parameter Value:
30 to 180 (The default is 60.)
Unit:
S
Response
Returns the d measurement time in the <NR3> format.
** RST
RST
** RST
RST
** RST
RST
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......