308
KHA3000
Harmonic Measurement Mode (61000-3-12)
Setting the Judgment Rsce
HARM:IEC312:JUDG:RSCE
Sets the judgment Rsce.
Command
HARMonics:IEC312:JUDGe:RSCE{<numeric>|MIN|MAX}
HARMonics:IEC312:JUDGe:RSCE?
Parameter Value:
33 to 350 (The default is 33.)
Response
Returns the judgment Rsce in the <NR1> format.
Setting the Limit Value to be Applied
HARM:IEC312:LVT
Sets the limit value to be applied.
Command
HARMonics:IEC312:LVTable {UNBalanced}|BALanced|SPECified}
HARMonics:IEC312:LVTable?
Parameter Value:
UNBalanced
Except balanced three-phase equipment (default)
BALanced
Balanced three-phase equipment
SPECified
Specified three-phase equipment
Response
Returns the limit value to be applied in the <character> format.
Setting the Nominal System Voltage
HARM:IEC312:NOM:SVOL
Sets the nominal system voltage.
Command
HARMonics:IEC312:NOMinal:SVOLtage <NRf>
HARMonics:IEC312:NOMinal:SVOLtage?
Parameter Value:
100 to 600 (The default is 400.)
Unit:
V
Response
Returns the nominal system voltage in the <NR3> format.
** RST
RST
** RST
RST
** RST
RST
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......