292
KHA3000
Measurements in General
Setting the Operation Mode
FUNC
Sets the operation mode.
Command
[SENSe:]FUNCtion[:ON]
“{HARMonics|VFL|OTHer:BASic|OTHer:FFT|OTHer:RUSH}”
[SENSe:]FUNCtion[:ON]?
Parameter Value:
“HARMonic
s
”
Harmonic measurement mode (default)
“VF
L
”
Voltage fluctuation measurement mode
“OTHer:BASic”
Basic measurement
“OTHer:FFT”
FFT analyzer
“OTHer:RUSH”
In-rush current measurement
Response
Returns the present operation mode in the <string> format.
Returns OTH when the top page of Other Measurements is displayed.
Returns
“
”
(blank) when the System Setup or External Control page is displayed.
Setting Linked/independent of the Input Circuits
INP:LINK
Sets the input circuit of each input phase (channel) to linked or independent.
Command
INPut:LINKage {ON|OFF|1|0}
INPut:LINKage?
Parameter Value:
ON(1)
linked (default)
OFF(0)
independent
Response
Returns linked or independent in the <NR1> format.
Setting the Current Input Terminals
INP:CURR:TERM
Sets the current input terminals. Each input phase (channel) is set to the same value. This
command is enabled when the input circuit setting is
“
linked
”
(INP:LINK ON).
Command
INPut:CURRent:TERMinal {SHUNt|CLAMp|BNC}
INPut:CURRent:TERMinal?
** RST
RST
** RST
RST
** RST
RST
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......