Ex
ecuting V
olta
g
e Fluctuation T
ests
When the Test Time Expires (61000-3-3)
●
When the result of the general judgment is acceptable (PASS,the progress bar is green),
follow the procedure described below.
●
When the result of the general judgment is not acceptable (FAIL,the progress bar is red),
set the impedance again and re-execute the test for the 61000-3-11 standard. For details,
see "Setting the Impedance Again (61000-3-11)".
When the General Judgment is PASS
1
Press the F1 key (OK).
The test ending menu shown below is displayed and the buzzer stops.
2
When terminating the test, select an item you want to operate from the
test ending menu.
For details on the selection, see "Overview of the Test Ending Menu".
When the General Judgment is FAIL
�
Set the impedance again and re-execute the test for the 61000-3-11
standard.
See "Setting the Impedance Again (61000-3-11)".
S
ee
Display after the test ends
Test status display
Switches from
“
Test
”
to
“
Analysis.
”
The progress bar extends to the right end and its color changes
according to the judgment.
PASS
The result of the general judgment is acceptable (the progress bar is
green).
WARNING
The result of the general judgment is acceptable but the margin is
exceeded (the progress bar is yellow).
FAIL
The result of the general judgment is not acceptable (the progress bar is
red).
S
ee
S
ee
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......