166
KHA3000
View Setting Sub-menu
3D Harmonics
•
Display color changes.
•
Used for observation and analysis.
Meter
No.
Menu item
Description
Default
1
Vertical Scale (Current)
Select a vertical scale of current (scale factor).
The vertical scale is displayed in the following manner. "log" denotes
logarithmic scale.
61000-3-2: Current
61000-3-12 2011: Percentage of the rms current
*1 ,*2
61000-3-12 2004: Percentage of the fundamental current
*2
*1.
In 61000-3-12 2011, in place of displaying the percentage of the reference current Iref, which is the rms current
during the test period, the instantaneous rms current is used.
*2.
The bar graph of the 1st harmonic order is not displayed.
×
1
2
Check View Phase
Select a phase to be displayed.
L1
3
Return
Terminates the view setting.
–
F key: Function key. You can also use the ESC key to return.
Graph
Judgment
Red
Part exceeding 100% of the limit value
Yellow
Part exceeding the set margin value and up to 100% of the limit value
Green
Part up to the set margin value
F1
F5
L1
L2
L3
F6
Return
1
3
2
10,
5,
2,
1,
1/2,
1/4, log
Vertical Scale (Current)
Check View Phase
F1
Peak Reset
F5
Check View Phase
L1
L2
L3
F6
Return
1
2
3
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......