8-20
KHA1000
8.7
SCPI Commands Used for This Product
8.7.1
Measurements in General
FUNC
Selects the operation mode.
Command
[SENSe:]FUNCtion[:ON]
”{HARMonic|VF|OTHer:BASic|OTHer:FFT|OTHer:RUSH}”
[SENSe:]FUNCtion[:ON]?
Parameter
Value:
“HARMonic”
Harmonic current test (Default)
“VF”
Voltage fluctuation test
“OTHer:BASic”
Basic measurement
“OTHer:FFT”
FFT analyzer
“OTHer:RUSH”
In-rush current measurement
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the present operation mode as character string data.
Returns OTH when the top page of other measurements is displayed.
Returns “ ” (blank) when the system setup or EXT control page is displayed.
CURR:COUP
VOLT:COUP
Sets the input coupling. Executing either command to set a value automatically sets
the other command to the same value.
Command
[SENSe:]CURRent:COUPling {AC|DC}
[SENSe:]VOLTage:COUPling {AC|DC}
[SENSe:]CURRent:COUPling?
[SENSe:]VOLTage:COUPling?
Parameter
Value:
AC
Alternate current coupling
DC
Direct current coupling (Default)
When *RST or MEAS:<meter_fn>? is sent, settings are made as shown in Table 8-4.
Response
Returns the setting for the input coupling.
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......