Select an estimated maximum value in the operation cycle of the EUT. The peak
current that can be measured is four times the value in the 0.5 A to 10 A range and
2.5 times the value in the 20 A range.
•
The maximum value of the input current is 50 Apeak. Exceeding this
value may cause overheating of the current detector.
•
If the current detector overheats, the OHP icon is displayed in the upper
part of the window. In this case, immediately shut off the power to the
EUT to cut input current of this product. Restart the test after the OHP
icon disappears.
The peak current of the EUT is believed to be reached at maximum power. Keep in
mind the entire operation cycle of the EUT.
Nominal values (general values) of test voltage and frequency. Select them
according to the rating of the EUT.
Nominal value
Sub Menu
CAUTION
F5
Nominal Values
Sub Menu
Nominal
Voltage
Nominal
Frequency
F1
Nominal Voltage
100 V
120 V
200 V
230 V
F2
Nominal
Frequency
50 Hz
60 Hz
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......