5.1
Setting IEC 61000-3-2 (Ed2.2/Ed3.0/Ed5.0) Test
Conditions
Set test conditions in the HA Test Conditions List display. The conditions thus set
are also used for the HA Observation and Test Conditions display.
Fig.5-1
HA Test Conditions List display IEC Ed2.2/Ed2.0
Showing the HA Test Conditions List Display
Press the HA key.
The HA key LED illuminates, and the HA Test Conditions List display appears.
While the test status shown on the upper left of the display
is “Test” or “Analysis,”
the mode cannot be changed. The dialog box message “Can’t execute during test/
analysis. Please operate it after ending” is displayed.
In this case, press the VIEW key to turn off its LED. Select “Exit” from the
displayed menu.
Test conditions can be set during measurement.
Set test conditions in the HA Observation and Test Conditions display. The
measurement display is set in the HA Observation and Analysis display (HA-
VIEW) and uses the same menu as in the HA Test Conditions List display.
View selection and transition
1.
Press the HA key.
The HA key LED illuminates and the HA Test Conditions List display appears.
2.
Press the VIEW key.
The VIEW key LED illuminates, and the HA Observation and Analysis display (HA-
VIEW) appears.
F1
F2
F3
F4
F5
F6
See
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......