4-44
KFM2151
Remote Control
The MEAS:VOLT? allows you to easily make measurements even if you do not
have a detailed knowledge of the KFM2151. The down side is that all the settings
related to the measurement functions are set to default. This prohibits you from
setting detailed measurement conditions. To make measurements using setup
conditions with a little more detail, use the CONF:VOLT command and READ?
query.
The MEAS:VOLT? query is equivalent to combining the CONF:VOLT command
and READ? query.
4.9.2
Measurement procedure with a little more flexibility
(CONFigure and READ)
The measurement procedure using the CONF:VOLT command and READ? query
separates the measurement condition setup from the measurement start and data
acquisition procedures. This enables you to specify detailed conditions between the
setup and measurement start procedures. For the default settings set by the
CONF:VOLT command, see Table 4-20.
You can set the settings to default using the CONF:VOLT command, specify other
settings, and then start the measurement using the READ? query.
Example:
CHAN:VOLT:NSEL 1
' Select logical
channel 1.
CONF:VOLT
' Set voltage scanning
to default.
SENS:VOLT:RANG 20V
' Set the selected
channel to 20 V range
SENS:VOLT:PROT:UPP 16;LOW 2
' Set the selected
channel to OVP 16 V and
UVP 2 V.
READ?
' Start voltage
scanning on all
activated channels, and
query the measured
data.
(Read the response)
To abort the measurement after it is started, use the ABORt command. You can also
use the *RST command or break signal to abort the measurement. The *CLS
command does not abort the measurement. Once the measurement starts, you
cannot enter settings related to the test conditions. You can specify the settings again
after the test is completed or aborted.
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......