4-20
KFM2151
Remote Control
4.5.4
Configuration Function
CONF:VOLT
Sets the voltage range of the channel specified by CHAN:VOLT:NSEL and sets the
voltage protection (OVP and UVP) to default.
Command
CONFigure[:SCALar]:VOLTage[:DC]
[{<numeric>|MIN|MAX|DEF}]
Parameter
<numeric>Voltage range setting
V
al
u
e:
2, 20, or 200 The default value is AUTO.
This command is the same as sending the following commands.
[SENSe:]VOLTage[:DC]:RANGe <numeric>
[SENSe:]VOLTage[:DC]:PROTection[:UPPer] DEFault
[SENSe:]VOLTage[:DC]:PROTection:LOWer DEFault
CONF:VOLT:ALL
Sets the voltage range of all logical channels that are set as scanned sources for
voltage scanning and sets the voltage protection (OVP and UVP) to default.
Command
CONFigure[:SCALar]:VOLTage[:DC]:ALL
[{<numeric>|MIN|MAX|DEF}]
Parameter
<numeric>Voltage scanning range setting
V
al
u
e:
2, 20, or 200 The default value is AUTO.
This command is the same as sending the following commands to all channels set as
scanned sources.
[SENSe:]VOLTage[:DC]:RANGe <numeric>
[SENSe:]VOLTage[:DC]:PROTection[:UPPer] DEFault
[SENSe:]VOLTage[:DC]:PROTection:LOWer DEFault