KFM2151
3-13
Operation
3.8
Channel Expansion
Up to four slave units can be connected to a single master unit.
The number of channels can be expanded to 160 by using five KFM2151s.
For the connection between the KFM2151 and the KFM2150 FC Impedance Meter,
see “2.4 Connecting the KFM2151 to the FC Impedance Meter.”
Table 3-5 shows the channel numbers when multiple KFM2151s are used.
If you are using a single KFM2151, set it to be the master unit.
Channel verification works correctly only when the master unit is operating.
Ta
b
le 3-5
Channel n
u
m
b
ers of the master
u
nit and sla
v
e
u
nits
Setting the master unit or slave unit (unit number)
Set one KFM2151 to be the master unit (unit number: 0) and other KFM2151s to be
slave units (unit number: 1 to 4).
1.
Select parameter n
u
m
b
er 010 in the CO
N
FIG settings.
2.
In the condition setting, set the master
u
nit to 00 and sla
v
e
u
nits to 01,
02, 03, or 04.
3.
T
u
rn the PO
W
ER s
w
itch off and then t
u
rn it
b
ack on.
See
Master/Sla
v
e
Unit
Nu
m
b
er
Channel
Nu
m
b
er
Master
u
nit
(factory defa
u
lt)
0
001 to 032
Sla
v
e
u
nit
1
101 to 132
2
201 to 232
3
301 to 332
4
401 to 432
See
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......