5-4
KFM2151
Maintenance
■
Does not operate according to the settings.
Symptom
Check Items
Remedy
Page
The settings entered
from the CO
N
FIG are
not applied.
• Did yo
u
t
u
rn off the PO
W
ER
s
w
itch after changing the
settings?
• T
u
rn off the PO
W
ER s
w
itch
after yo
u
change the settings.
The settings are applied
w
hen the po
w
er is t
u
rned
b
ack on.
Settings are not
applied.
• Did yo
u
press the SET key
w
hen yo
u
w
ere done
w
ith the
settings?
• Enter the settings again and
press the SET key.
The O
V
P LED is
ill
u
minated in red.
• Is the O
V
P for the channel on
w
hich the monitor LED is
ill
u
minated in red set
appropriately?
• Is the f
u
el cell
v
oltage high?
• The o
v
er
v
oltage protection
(O
V
P) is acti
v
ated. Check
that the O
V
P for the channel
on
w
hich the monitor LED is
ill
u
minated in red set
appropriately or check the
f
u
el cell.
The U
V
P LED is
ill
u
minated in red.
• Is the U
V
P for the channel on
w
hich the monitor LED is
ill
u
minated in red set
appropriately?
• Is the f
u
el cell
v
oltage lo
w
?
• The
u
nder
v
oltage protection
(U
V
P) is acti
v
ated. Check
that the U
V
P for the channel
on
w
hich the monitor LED is
ill
u
minated in red set
appropriately or check the
f
u
el cell.
See
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......