46
DSI1020/ DSI3020
Specifications
Voltage-dip and short-interruption test performance
Compliant standards
IEC 61000-4-11 Ed.2.0:2004
Phase
1P2W
• Model DSI1020
1P2W, 1P3W,
3P3W, 3P4W
• Model DSI3020
Dip source
Phase voltage
• Model DSI1020
Line voltage, phase voltage
• Model DSI3020
Voltage dips
voltage setting accuracy
1
1
The specifications are relative to the phase-voltage variation.
±5 %
Voltage dips
voltage load regulation
1
< 5 % of the U
T
2
2
U
T
denotes the nominal test voltage.
Maximum input voltage
1
288 Vrms
• Phase voltage (between U, V, W and N)
500 Vrms
700 Vpeak
• Line-voltage (U-V, V-W, and W-U)
• Model DSI3020
Maximum output
current
Rms
value
16 Arms/phase
• At test voltage 100 %U
T
40 Arms/phase
• At test voltage 40 %U
T
Peak
value
500 Apeak
• 1 s or less
Voltage overshoot
1
< 5 % of the U
T
• Test voltage U
T
100 V to 240 V, load 100
Ω
When varied from 0 %U
T
to 100 %U
T
(phase angle 90°)
Voltage undershoot
1
< 5 % of the U
T
• Test voltage U
T
100 V to 240 V, load 100
Ω
When varied from 100 %U
T
to 0 %U
T
(phase angle 90°)
Voltage rise time
1
1
μ
s to 5
μ
s
• Test voltage U
T
100 V to 240 V, load 100
Ω
When varied from 0 %U
T
to 100 %U
T
(phase angle 90°)
Voltage fall time
1
1
μ
s to 5
μ
s
• Test voltage U
T
100 V to 240 V, load 100
Ω
When varied from 100 %U
T
to 0 %U
T
(phase angle 90°)
Voltage dips
phase setting accuracy
1
< 10°
• Test voltage U
T
100 V to 240 V, load 100
Ω
Phase angle 0°, 45°, 90°, 135°, 180°,
225°, 270°, 315°, 360°
Voltage dips
duration setting
0.5, 1, 5, 10, 25, 50 cycles, 2 to 300 cycles
(can be set to any cycle value. Resolution 1 cycle)
Interval time
setting accuracy
±10 %
• Setting range 10 s to 99 s (default 10 s)