62
User’s Manual
BIM1000 Series
Remote Control | Message Reference
Queries the status
Queries the condition of the QUEStionable status register.
A query does not clear the contents of the register.
QUEStionable status enable register
Sets or queries the QUEStionable status enable register. Set using a total value of the weight
bits you want to set.
Switching the measurement display
Switches the measured value that is shown on the front panel
Resistance measurement range
When you send a resistance value you want to measure, the range is set appropriately for measuring that
resistance.
Program header
:STATus:QUEStionable:CONDition?
Value/Response
<0 to 3845 (NR1)>
Program header
:STATus:QUEStionable:ENABle
:STATus:QUEStionable:ENABle?
Value/Response
<0 to 3845 (NR1)>
Example
:STATus:QUEStionable:ENABle 1536
Set QUEStionable status enable register bit 9 (bit weight 512) and bit 10 (bit weight
1024) to 1.
Program header
:FUNCtion
:FUNCtion?
Value
{RV|RESistance|VOLTage}
RV:
Displays the resistance and voltage
RESistance:
Displays the resistance
VOLTage:
Displays the voltage
Response
{RV|RESISTANCE|VOLTAGE}
Program header
:RESistance:RANGe
:RESistance:RANGe?
Value
<0 to 3.1 (NR2)>
Unit
Ω
Response
<NR3>
Example
Send “: RES: RANG 0.12”
The range is set to 300 mΩ, which allows 120 mΩ to be measured.
Send “: RES: RANG?”
The present range setting “3.00000E-01” (300 mΩ) is returned.
Note
If :AUTorange is set to ON (1) or RES, the value changes automatically.