Remote Control | Registers
Event status register
This register notifies the status byte register of the product’s status. There are three types of event status
registers: standard event status register, event status register 0, and event status register 1. When the
product’s status changes, the corresponding bit in the event status register is set to 1. If the bit set to 1 is
enabled, the corresponding bit (ESB, ESB0, ESB1) in the status byte register is set to 1.
To enable the bits in the event status byte register, use the following commands to set the corresponding
bits in the enable register.
Standard event status enable register: *ESE
Event status enable register 0: :ESE0
Event status enable register 1: :ESE1
Standard event status register (SESR)
&
&
&
&
&
&
ESB
ESB0
ESB1
&
&
&
&
&
&
&
&
Event status
register 0
EOM
INDEX
NOT USED
NOT USED
ERR
NOT USED
NOT USED
NOT USED
R-Low
R-IN
R-Up
V-Low
V-Up
PASS
V-IN
FAIL
1
0
2
3
4
5
6
7
1
0
2
3
4
5
6
7
Event status
register 1
Event status
enable register 1
OPC
OPC
NOT USED
QYE
DDE
CME
NOT USED
EXE
PON
NOT USED
QYE
DDE
CME
NOT USED
EXE
PON
1
0
2
3
4
5
6
7
1
0
2
3
4
5
6
7
Standard event
status register
Standard event
status enable register
Event status
enable register 0
EOM
INDEX
NOT USED
NOT USED
ERR
NOT USED
NOT USED
NOT USED
1
0
2
3
4
5
6
7
1
0
2
3
4
5
6
7
&
&
&
Status byte register
R-Low
R-IN
R-Up
V-Low
V-Up
PASS
V-IN
FAIL
1
0
・
・
・
5
・・
Bit
Bit
weight
Bit
name
Description
0
1
OPC
Set when an *OPC command is received and all operations in standby have
been completed.
1
2
–
Not used
2
4
QYE
Set when an attempt is made to read data from the output queue when there is
no data or when the output queue is not in the wait state. This indicates that there
is no data in the output queue.
3
8
DDE
Set when there is a device-specific error.
4
16
EXE
Set when the product evaluates that the program data after the header is outside
the formal input range or does not match the specifications of the product. This
indicates that a valid SCPI command may not be executed correctly depending
on the state of the product.
5
32
CME
Set when an IEEE 488.2 syntax error is detected by the parser, when an uniden-
tifiable header is received, or when a group execution trigger enters the internal
IEEE 488.2 SCPI command input buffer.
6
64
–
Not used
7
128
PON
Set when the power is turned on.