System Settings | General Settings
Measurement end signal output (EOM)
Set how to output the measurement end signal (EOM). EOM is output from the SIGNAL I/O connector. For
timing charts, see “When starting measurements with internal triggers”
ments with external trigger input”
.
1
Press SYSTEM.
2
Press EOM (F5) to select the value.
The value switches each time you press the F5 key.
3
If you specify PULSE, enter the time using the numeric keypad, and then press
ENTER.
This completes the setting.
Value
Description
HOLD
is set to INTERNAL, the signal is turned on after a mea-
surement is completed until the next measurement starts. When the trigger source is set to
EXTERNAL, the signal is turned on after a measurement is completed until the next trigger
is detected.
PULSE
The signal turns on for the specified time when a measurement is completed.